bowie dick test

10 Common Causes for Failure of a Bowie-Dick Test

Amit Gupta
Written by: Amit Gupta

MS Mechanical Engineering, Vice President of Engineering

Key Takeaways

  • A Bowie-Dick test confirms that a prevacuum autoclave removes enough air for steam to penetrate wrapped goods and packs.
  • Bowie-Dick testing calls for a warm-up cycle, then a short cycle with the test pack alone on the lowest shelf above the drain.
  • Only a completely black thermochromic sheet indicates a passing test.
  • Common causes for failed tests include operator errors, incorrect cycle selection, air leaks, low vacuum levels, and non-condensable gas.
  • After a failed test, remove the autoclave from use until a repeat test passes.

Sterilization departments have one of the most important responsibilities within a healthcare facility: ensuring that all surgical instruments and other reusable equipment are sterilized and safe to use during upcoming procedures. Without these services, operations at hospitals and ambulatory surgery centers would grind to a halt.

Steam sterilization is equally important in laboratory settings (especially those that carry out infectious disease research), where it plays a key role in preventing the spread of pathogens and maintaining the integrity of samples and experiments.

So, whether you work in a lab or sterile processing department, it’s crucial to make sure that your autoclaves are always working properly. One way to do this is by running a Bowie-Dick test. Read on to find out more about this test, including how to conduct one and how to identify different failure cases.

What Is a Bowie-Dick Test?

A Bowie-Dick test is a standard operational test that hospitals and laboratories can use to determine whether their prevacuum is properly removing air from the chamber. It’s best for testing prevacuum cycles that are sterilizing wrapped goods or packs. Any pockets of cool air that remain in the chamber act as a barrier that prevents steam penetration into these packs, thus inhibiting sterilization.

The first test pack was developed in England by doctors J.H. Bowie and J. Dick in 1963 and, in the time since, has become the standard for confirming suitable autoclave performance. Today’s test packs include thermochromic paper to determine whether steam has suitably penetrated the porous load. A successful test shows a uniform, dark black color pattern on the paper and means that a successful vacuum and full steam penetration was achieved. No color change or a partial color change indicates a failed test.

How To Run a Bowie-Dick Test

Running a Bowie-Dick test requires placing a small, disposable test pack into the autoclave and running a four-minute sterilization cycle, typically a Bowie-Dick cycle. This sterilization cycle usually consists of three to four prevacuum pulses, each of which involve injecting steam into the autoclave and then pulling out the air and steam through a vacuum, before reaching the set point of 270°F (132°C).

Taking a closer look, the process can be further broken down into the following steps:

  • Perform a warm-up cycle. This step ensures that the autoclave is sufficiently heated up, which helps prevent false-fail test results.
  • Confirm that the autoclave chamber is empty. Bowie-Dick tests are always run in an otherwise empty autoclave. This makes it easier to detect air leaks or problems with its air-removal system as the presence of additional packs reduces the sensitivity of the test.
  • Place the test pack on the lowest shelf above the drain. The pack must be placed in this specific location because the drain is where the vacuum is drawn, putting it in the natural path of any residual air exiting the chamber.
  • Begin the Bowie-Dick test cycle. While parameters for these cycles aren’t specifically defined in the ANSI/AAMI ST79, general guidance recommends three-and-a-half to four minutes of exposure time. However, many modern autoclaves have pre-programmed cycles that are specifically designed for running Bowie-Dick tests.
  • Examine the test sheet and interpret results. Once the test cycle is complete, open the pack to see the results.

Autoclave expert Bruce Gillingham believes it is an imperative process that should be performed each and every day within a hospital setting. This is because many hospitals (and labs) end up shutting down their autoclave at the end of the day, which can lead to unexpected issues when it is turned on again the next day.

“We recommend running a routine five-minute sterilization cycle at the start of every day,” says Gillingham. “The initial five-minute cycle helps to get rid of any air pockets that might be in the jacket. Once this is complete, we recommend running a Bowie-Dick test before you sterilize your first load of the day.”

Interpreting the Test Results

According to Gillingham, “It’s the color change that tells you the steam is penetrating the test pack and pulling all the air out of the autoclave.”

A successful Bowie-Dick test is very evident based on the color change of the thermochromic paper within the test pack. If the thermochromic paper turns completely black, then the steam has penetrated the load and the autoclave is operating correctly.

Lack of color change or a partial color change indicates an unsuccessful Bowie-Dick test cycle. “If the color hasn’t completely changed, it means something hasn’t worked correctly,” says Gillingham.

bowie dick test

10 Common Causes for Bowie-Dick Test Failure

Whether it’s because of problems with the autoclaving unit, the test pack, or a result of human error, there are several explanations for why a Bowie-Dick test might produce a failed result. These are ten of the most common:

  1. Overloaded chamber: Bowie-Dick tests are designed to be run in an otherwise empty chamber, and additional items can change how air and steam move around the test pack and skew the result.
  2. Incorrect test placement: The test pack needs to lie flat on the lowest shelf directly above the drain, where residual air leaves the chamber. A pack placed elsewhere or on its side might not accurately reflect how well the autoclave removes air.
  3. Air leak: If air is able to leak into the autoclave chamber, the steam will be unable to penetrate the load to the point of total sterilization.
  4. Unwanted condensation: Occasionally, condensation gets trapped in the jacket of the autoclave, which can lead to cold spots at the base of the unit. This could also indicate a wet-steam issue.
  5. Faulty test pack: From time to time, a Bowie-Dick test pack can be faulty.
  6. No warm-up cycle: A warm-up cycle enables the sterilizer chamber and jacket to reach temperature; not running a warm-up cycle can cause a failure.
  7. Incorrect procedures: Test packs work under very specific conditions, so it’s important to make sure all steps are followed exactly when running the test.
  8. Low vacuum level: For a sterilization cycle to run successfully, an autoclave must create sufficient vacuum conditions. If the unit is incapable of producing these conditions, sterilization will not be achieved.
  9. Steam quality issues: Steam with excess moisture can soak the test pack and prevent complete penetration, and superheated steam does not condense as readily as saturated steam. Either condition can leave an uneven color change on the test sheet.
  10. Presence of non-condensable gas: Non-condensable gasses decrease sterilization efficacy and can make their way into the chamber via pneumatic valve leaks or water that has elevated levels of bicarbonate.

The good news? There are measures you can take to prevent these causes from occurring. Take a look at the following chart for some possible solutions.

Cause of Failure Possible Solution
Overloaded chamber Remove everything from the chamber except the test pack before you start the cycle.
Incorrect test placement With its label facing up, lay the pack flat on the bottom shelf, directly over the drain.
Air leak Run a vacuum leak test to further determine if an air leak exists or not.
Unwanted condensation Check the steam traps on the autoclave. Check steam quality and wetness.
Faulty test pack Check the expiration date and make sure the packs are being stored in the proper environment.
No warm-up cycle Run a 5 minute sterilization cycle prior to running the Bowie-Dick cycle.
Incorrect procedures Test packs should be placed in an empty chamber directly over the drain on the bottom rack or shelf. They are designed for use at 270°-273°F (132°-134°C).
Low vacuum level Check the vacuum pump and pressure switch, as either (or both) may be faulty. Another culprit could be the vacuum pump’s water supply — water that’s too hot may cause the pump to malfunction.
Poor steam quality Schedule steam quality testing and inspect the steam traps on the autoclave.
Presence of non-condensable gas If a leaky pneumatic valve is the issue, it will need to be repaired or replaced. Steam quality testing may also be in order, especially if you rely on a softened water source.

Next Steps after a Bowie-Dick Test

A successful Bowie-Dick test means you can begin sterilizing loads with a prevacuum cycle. Meanwhile, a failed Bowie-Dick test means further testing is required. If none of the above solutions (see table) work then both the autoclave and the utilities should be checked. The autoclave should not be used with any confidence until the test is repeated with a passing result.

Please contact us with any questions or if you would like more information about running Bowie-Dick tests or interpreting the results. If you’re looking for test packs for your sterilizers, you can also visit our autoclave supply shop to stock up.

Frequently Asked Questions

Can you run a Bowie-Dick test in a gravity displacement autoclave?

A Bowie-Dick test isn’t meaningful in a gravity displacement autoclave because it’s meant to evaluate mechanical air removal, which gravity units do not do. In a gravity cycle, steam pushes air out through a drain with no vacuum assistance, so the test has no vacuum system to assess. Facilities with gravity autoclaves can monitor their performance with chemical indicators, biological indicators, and cycle records, instead.

What’s the difference between a Bowie-Dick test and a biological indicator?

A Bowie-Dick test is a daily check of a prevacuum autoclave’s air removal, while a biological indicator measures whether a cycle killed highly heat-resistant bacterial spores.

What’s the difference between a Bowie-Dick test and a vacuum leak test?

A Bowie-Dick test shows whether a prevacuum autoclave removes enough air for steam to penetrate a porous load, whereas a vacuum leak test measures how much vacuum the chamber and its plumbing lose over a set period. A leak test can detect a slow leak before it grows large enough to cause a Bowie-Dick failure, so the two tests complement each other.

Does a passing Bowie-Dick test mean a load is sterile?

A passing Bowie-Dick test does not mean a load is sterile because the test only confirms that the autoclave removed air from an empty chamber.

What should you do with loads processed after a failed Bowie-Dick test?

Loads processed after a failed Bowie-Dick test should be quarantined and treated as nonsterile, and any that have already been released should be recalled. The recall should cover every load run in that sterilizer since its last passing test. Once the autoclave is serviced and passes a repeat test, the recalled items can be repackaged and reprocessed.

Do you need Bowie-Dick test packs?

Get your Bowie-Dick test packs at the CSS store now.

Visit Store
Do you need Bowie-Dick test packs?